IMIS

Publicaties | Instituten | Personen | Datasets | Projecten | Kaarten
[ meld een fout in dit record ]mandje (0): toevoegen | toon Print deze pagina

Comparison of multiple methods for quantification of microbial biofilms grown in microtiter plates
Peeters, E.; Nelis, H.J.; Coenye, T. (2008). Comparison of multiple methods for quantification of microbial biofilms grown in microtiter plates. J. microbiol. methods 72(2): 157-165. https://dx.doi.org/10.1016/j.mimet.2007.11.010
In: Journal of Microbiological Methods. Elsevier: Amsterdam. ISSN 0167-7012; e-ISSN 1872-8359, meer
Peer reviewed article  

Beschikbaar in  Auteurs 

Author keywords
    Biofilm quantification; Microtiter plate; Crystal violet; Resazurine; FDA; XTT

Auteurs  Top 
  • Peeters, E.
  • Nelis, H.J.
  • Coenye, T., meer

Abstract
    In the present study six assays for the quantification of biofilms formed in 96-well microtiter plates were optimised and evaluated: the crystal violet (CV) assay, the Syto9 assay, the fluorescein diacetate (FDA) assay, the resazurin assay, the XTT assay and the dimethyl methylene blue (DMMB) assay. Pseudomonas aeruginosa, Burkholderia cenocepacia, Staphylococcus aureus, Propionibacterium acnes and Candida albicans were used as test organisms. In general, these assays showed a broad applicability and a high repeatability for most isolates. In addition, the estimated numbers of CFUs present in the biofilms show limited variations between the different assays. Nevertheless, our data show that some assays are less suitable for the quantification of biofilms of particular isolates (e.g. the CV assay for P. aeruginosa).

Alle informatie in het Integrated Marine Information System (IMIS) valt onder het VLIZ Privacy beleid Top | Auteurs