IMIS

Publications | Institutes | Persons | Datasets | Projects | Maps
[ report an error in this record ]basket (0): add | show Print this page

Determination of cadmium in marine waters: on-line preconcentration and flow-through fluorescence detection
Charles, S.; Yunus, S.; Dubois, F.; Vander Donckt, E. (2001). Determination of cadmium in marine waters: on-line preconcentration and flow-through fluorescence detection. Anal. Chim. Acta 440(1): 37-43. http://dx.doi.org/10.1016/S0003-2670(01)00857-1
In: Analytica Chimica Acta. Elsevier: New York; Amsterdam. ISSN 0003-2670; e-ISSN 1873-4324, more
Peer reviewed article  

Available in  Authors 

Keyword
    Marine/Coastal
Author keywords
    Cadmium; On-line preconcentration; Molecular fluorescence spectroscopy; Seawater; Anthrylazamacrocycle; Fluorogenic reagent

Authors  Top 
  • Charles, S.
  • Yunus, S.
  • Dubois, F.
  • Vander Donckt, E.

Abstract
    This paper reports the development of a new molecular fluorescence spectroscopy method for the determination of cadmium in marine waters. An anthrylhexaazamacrocycle was used as a fluorogenic reagent at pH 13. Ionic extraction chromatography on Dowex resin was necessary to separate cadmium from the matrix and especially from lead, mercury and organic matters present in seawater and interfering with the fluorescence measurements. The coupling of on-line preconcentration by Dowex resin and fluorescence flow-through detection at pH 13 led to a detection limit of 35 pM for a preconcentration time of 5 min (2.1 ml min-1). The accuracy of the method was verified on certified seawater samples SLEW-2 and NASS-5. Water samples collected in the Scheldt estuary near Antwerp (Belgium) during an ebb tide were analysed for dissolved cadmium.

All data in the Integrated Marine Information System (IMIS) is subject to the VLIZ privacy policy Top | Authors